Per section 807.81 - When a premarket notification submission is required.

(a) Except as provided in paragraph (b) of this section, each person who is required to register his establishment pursuant to 807.20 must submit an FDA 510K premarket notification submission to the Food and Drug Administration at least 90 days before he proposes to begin the introduction or delivery for introduction into interstate commerce for commercial distribution of a device intended for human use which meets any of the following criteria:

(1) The device is being introduced into commercial distribution for the first time; that is, the device is not of the same type as, or is not substantially equivalent to, (i) a device in commercial distribution before May 28, 1976, or (ii) a device introduced for commercial distribution after May 28, 1976, that has subsequently been reclassified into class I or II.

(2) The device is being introduced into commercial distribution for the first time by a person required to register, whether or not the device meets the criteria in paragraph (a)(1) of this section.

(3) The device is one that the person currently has in commercial distribution or is reintroducing into commercial distribution, but that is about to be significantly changed or modified in design, components, method of manufacture, or intended use. The following constitute significant changes or modifications that require a premarket notification:

(i) A change or modification in the device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process.

(ii) A major change or modification in the intended use of the device.

(b) A premarket notification under this subpart is not required for a device for which a premarket approval application under section 515 of the act, or for which a petition to reclassify under section 513(f)(2) of the act, is pending before the Food and Drug Administration.

(c) In addition to complying with the requirements of this part, owners or operators of device establishments that manufacture radiation-emitting electronic products, as defined in 1000.3 of this chapter, shall comply with the reporting requirements of part 1002 of this chapter.

Email Us

or

Call us toll free: 1-888-299-0157


• Home


FDA Compliance Help Desk™

1289 N. Fordham Blvd. Suite A-128

Chapel Hill, NC 27517 USA

Phone 919.942.1634 

info@FDAcomplianceHelp.com

Copyright © 2004 - 2005 FDA Compliance Help Desk, Inc.
Last modified: October 16, 2005

FDA Compliance Help Desk, Inc. is a for profit NC corporation and is not affiliated with the U.S. Department of Health and Human Services.